I s s u e 6 ( F e b r u a r y 2 0 0 5 )
Technical Ser vice Manual
S m i t h s M e d i c a l
4 — 8
4
M S 1 6 A S y r i n g e D r i v e r
iii.
Apply a smear of silicone grease (Smiths Medical
part number 6835-2006
) to the edge of the slotted head screw
(items 11 and 12).
b.
Assemble the
START/TEST
pushbutton, seal and switch bar
(items 5, 6 and 7) and place them in position over microswitches
SW1 and SW2.
Ensure that the switch bar is aligned across the contacts of SW1
and SW2.
Ensure that the cam lever (part of item 15) is on SW3 and that it
pivots freely.
c.
Ensure that the foam pad is fitted to the inside of the front case
(see
Figure 4-4, Foam pad position
). Place the front case half
(item 1) over the PCB and hold it in position. Turn the PCB and
case over so that the PCB is uppermost. Check that the leads for
the motor and battery connector springs are not trapped.
Figure 4-4 Foam pad position
d. Push the battery connector springs into their slots. Red to + and
Black to - ve, as shown on the battery compartment label.
e.
Secure the PCB in position with five screws and fibre washers.
The switch pin insulator (item 24) must be held by the lower
middle screw and separate the PCB from the piezo-transducer.
f.
Check for correct operation of the
START/TEST
pushbutton and
that clockwise rotation of both rate switches causes the dials to
indicate increasing numbers.
g.
Place the motor and gearbox in position and fit the insulating
spacer (item 23). Fit the motor clamp (item 13) with the securing
screw and washer; do
not
tighten the screw.
SM0105-4-6.PMD
2/23/2005, 2:56 PM
8