I s s u e 6 ( February 2 0 0 5 )
Technical Ser vice Manual
S m i t h s M e d i c a l
5 — 4
M S 1 6 A S y r i n g e D r i v e r
5
Adjustment potentiometers
The adjustment potentiometers VR5 and VR9 are behind the front
panel label.
Peel back the front panel label from the righthand side (nearest the
LED) to expose the access holes to the resistors.
Variable resistor VR9 (thrust) is located directly above the LED.
VR5 is near the top.
When adjustments have been completed, it may be necessary to
replace the front panel label if it does not adhere when pressed back
in position.
When replacing the label, ensure that the small clear plastic window
is in its recess over the rate dials.
Notes
: If any unit is found to be over or under infusing during
testing, inform your local RAQA Department.
Ensure that the button surround on the actuator body is blue
(
Figure 7-1, items 15 and 14
)
SM0105-5-6.PMD
2/23/2005, 3:07 PM