504-2056A 39
SECTION 8
TECHNICAL INFORMATION
(a) Principles
Of
Operation
The principle of operation of the babyPAC ventilator is illustrated by the block diagram in Fig 4,
page 42. Compressed oxygen and/or air enter the ventilator from the input hoses at the inlet
connectors
1
and
2
. The priority valve
3
supplies either oxygen or, if available, air to the internal
pressure regulator
4
where it is regulated to a constant pressure before being fed through to the
CMV valve
5
and the CPAP valve
23
.
When CMV is selected valve
5
is opened and gas passes through to the oscillator
6
.
The oscillator is a miniature pneumatic device which switches on and off at time intervals
determined by the settings of the precision needle valves T
INS
7
and T
EXP
8
.
When the oscillator is ‘on’, gas passes to the entrainment gas mixing device
9
where the gas
mixture is made up to the selected oxygen concentration at a fixed output flow of 10 L/min. From
here the gas passes to the ventilator gas output port and hence to the inspiratory limb of the patient
circuit.
The pressure in the patient circuit is set by means of the exhalation valve
17
. This is controlled by
means of the precision needle valve
19
working in conjunction with bleed restrictor
21
. Valve
19
is
fed, during the inspiratory phase, from the pneumatic switch valve
18
which is in turn actuated by
the oscillator output.
At the end of the inspiratory phase the oscillator output switches ‘off’ and hence the output flow
ceases and the valve
18
switches to supply the precision needle valve
20
instead of valve
19
. Valve
20
, in conjunction with bleed
21
, sets the PEEP pressure level on valve
17
and so once valve 18 has
switched the patient pressure will drop down to the PEEP level and the patient will exhale.
During the inspiratory phase the flow amplifier valve
11
supplies oxygen to the mixer control valve
12
to ensure that the required oxygen flow is supplied to the mixer
9
. The mixer draws in the
required volume of fresh air through inlet
10
.
The pressure supplied to the patient is limited by the adjustable upper pressure limiting valve
13
and
is monitored by means of the manometer
14
. Gas discharged from the pressure limiting valve is
used to sound the audible alarm.
The function switching valves
5, 23
and
24
are all actuated by the rotation of the function selector
switch shown as
4
in Fig 1(a).
When CMV+ACTIVE PEEP is selected then both valve
5
and valve
23
are opened. When the
oscillator is in its ‘on’ state its output pressure switches valve
27
‘off’. When the oscillator switches
‘off’ valve
18
switches as above but valve
27
subsequently opens and provides a gas supply to the
gas mixing device
9
and the flow amplifier valve
11
, thereby maintaining patient circuit flow
throughout expiration.
When CPAP is selected, the CMV valve
5
is closed and CPAP valve
23
is opened. In this way the
oscillator is bypassed and gas pressure is supplied continuously to the gas mixer
9
. The non-return
valve prevents this pressure actuating valve
18
so the PEEP/CPAP valve remains energised whilst
CPAP is selected and the setting of this valve determines the CPAP level.
When IMV+CPAP is selected all three valves
5, 23
and
24
are open. Valve
5
allows the oscillator
to function but valve
24
connects it to an auxiliary reservoir
25
which increases the expiratory time
Summary of Contents for Pneupac babyPAC 100
Page 2: ...504 2056A 2 Page Intentionally Blank ...
Page 10: ...504 2056A 10 Page Intentionally Blank ...
Page 26: ...504 2056A 26 Page Intentionally Blank ...
Page 38: ...504 2056A 38 Page Intentionally Blank ...
Page 41: ...504 2056A 41 Figure 4 ...
Page 42: ...504 2056A 42 Page Intentionally Blank ...
Page 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Page 52: ...504 2056A 52 Page Intentionally Blank ...
Page 56: ...504 2056A 56 Page Intentionally Blank ...
Page 58: ...504 2056A 58 Page Intentionally Blank ...
Page 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...