504-2056A
60
Table 3:
Inspiration & Expiration accuracies
Setting factory
calibrated
(seconds)
in field service
(seconds)
4.0 E time
± 0.60
± 0.80
2.0 I & E time
± 0.30
± 0.40
1.0 I & E time
± 0.15
± 0.20
0.75 I & E time
± 0.11
± 0.15
0.50 I & E time
± 0.08
± 0.10
0.35 I & E time
± 0.05
± 0.07
0.25 I & E time
± 0.1
± 0.1
Table 4:
Frequency Accuracies
Setting
Frequency range using I & E
service limits
Inspiration Expiration
Nominal
Frequency
Maximum Minimum
2.00 4.00
10.00 12.50
8.33
1.00 2.00
20.00 25.00
16.67
0.75 2.00
21.82 27.27
18.18
0.50 1.00
40.00 50.00
33.33
0.35 0.75
54.55 68.18
45.45
0.25 0.50
80.00 109.09 63.16