504-2056A
59
Appendix B Calibration accuracies and deviations due to change in ambient conditions
Table 2:
Calibration accuracies and deviations due to change in ambient conditions
Deviations due to change in ambient conditions
Accuracies of calibrated controls
Parameter Ambient
temperature
+40
°
C
Ambient
temperature
-10
°
C
Ambient
pressure
110 kPa
Ambient
pressure
70 kPa
Factory calibrated to
within :-
Further deviation
before re-calibration:-
T
I
min
+0.02 seconds
+0.07 sec
-0.015 sec
+0.07 sec
See Tables 3 & 4 below See Tables 3 & 4 below
T
E
min
-0.02 sec
+0.01 sec
-0.015 sec
+0.07 sec
See Tables 3 & 4 below See Tables 3 & 4 below
T
I
max
0 sec
+0.22 sec
-0.025 sec
+0.19 sec
See Tables 3 & 4 below See Tables 3 & 4 below
T
E
max
-0.24 sec
+0.74 sec
-0.25 sec
+1.0 sec
See Tables 3 & 4 below See Tables 3 & 4 below
T
E
IMV max
-2.0 sec
+2.0 sec
-2.0 sec
+2.0 sec
±
7%
±
8%
Peak Pressure
±
2 x100Pa
- 3 x100Pa
-1.5 x100Pa
+7 x100Pa
±
7%
±
8%
PEEP
±
1 x100Pa
- 4 x100Pa
-1 x100Pa
+4 x100Pa
±
7%
±
8%
CPAP
- 1 x100Pa
- 2 x100Pa
-1 x100Pa
+4 x100Pa
±
7%
±
8%
VRV pressure
- 1 x100Pa
- 4 x100Pa
-3 x100Pa
0
±
7%
±
8%
Oxygen % (@80%)
0
0
- 2%
+ 7%
±
5%
±
3%
Inflation pressure monitor
- 1 x100Pa
-1.5 x100Pa
- 1 x100Pa
+ 1 x100Pa
±
2 x100Pa
±
8% of reading
Summary of Contents for Pneupac babyPAC 100
Page 2: ...504 2056A 2 Page Intentionally Blank ...
Page 10: ...504 2056A 10 Page Intentionally Blank ...
Page 26: ...504 2056A 26 Page Intentionally Blank ...
Page 38: ...504 2056A 38 Page Intentionally Blank ...
Page 41: ...504 2056A 41 Figure 4 ...
Page 42: ...504 2056A 42 Page Intentionally Blank ...
Page 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Page 52: ...504 2056A 52 Page Intentionally Blank ...
Page 56: ...504 2056A 56 Page Intentionally Blank ...
Page 58: ...504 2056A 58 Page Intentionally Blank ...
Page 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...