504-2056A 20
The patient connection port is a 15mm female conical connector.
22.
Gas Supply Hoses
The hoses supplied with the babyPAC ventilator are 2 meters long, 3mm bore and colour
coded to BS 1319 and ISO 32. They are classed as “permanently connected” and are not to
be used for any purpose other than supplying the babyPAC. The user connectable probes
are gas specific and conform to BS 5682. A range of alternative oxygen and air gas supply
hoses is available with connections, probes and colours in conformity to other Standards.
23.
Carry Sling Attachment Slots
The carrying sling, W7140, is attached to the control module by means of slots as shown in
Fig. 1b.
24.
Mounting Attachment Points
Six M5 x 1.0 x 7mm deep female threaded bushes on the back of the control module may be
used for the attachment of brackets or for direct mounting.
25. Audible
alarm
The audible high pressure alarm is pneumatically powered by the gas relieved by the upper
pressure limiting system and warns that the limiting pressure has been reached.
26. Single
Gas
Operations
A green visual indicator is provided to indicate that the unit is operating on one gas only
(oxygen or air). The green LED illuminates 3 times every 30 seconds whenever the
ventilator is operating on a single gas supply. If one of the gases fails the audible alarm will
also sound until muted (section 2 (c) item11).
(e) Mounting
Options
The babyPAC ventilator has been designed to be mounted and carried with a wide range of options,
examples of which are shown in Fig. 2.
It may be mounted with its base or back on a flat surface. Alternatively it can be carried on the
shoulder by means of the carrying sling as shown in Fig. 2c.
For more permanent installations babyPAC ventilators can be fitted with a rail or pole mounting
bracket as shown in Fig 2 (d) and 2 (e). The brackets are attached to the back of the control
module by means of two M5 x 1.0 socket head fixing screws supplied with the bracket. The rail
bracket is of the universal type and attaches to 35/30 mm rails of either 6.5 or 10 mm thickness and
of the form shown in Fig 2 (d). A quick release clamp with a screw back-up is used.
The pole mounting bracket attaches to the module in the same way and accommodates vertical pole
diameters of 1/2 - 1" (12 - 26 mm). A screw with a T-handle is used for clamping.
The babyPAC ventilator can also be used mounted on transport incubators. Mounting kits are
available for certain widely used models. Contact Pneupac
®
for further information.
Summary of Contents for Pneupac babyPAC 100
Page 2: ...504 2056A 2 Page Intentionally Blank ...
Page 10: ...504 2056A 10 Page Intentionally Blank ...
Page 26: ...504 2056A 26 Page Intentionally Blank ...
Page 38: ...504 2056A 38 Page Intentionally Blank ...
Page 41: ...504 2056A 41 Figure 4 ...
Page 42: ...504 2056A 42 Page Intentionally Blank ...
Page 51: ...504 2056A 51 APPENDIX A Product Safety Transportation and Disposal of Recommended Batteries ...
Page 52: ...504 2056A 52 Page Intentionally Blank ...
Page 56: ...504 2056A 56 Page Intentionally Blank ...
Page 58: ...504 2056A 58 Page Intentionally Blank ...
Page 61: ...504 2056A 61 APPENDIX C Cleaning and inspection ...