System Fault Alarm
An unrecoverable error may
have occurred, such as a
hardware or software fault.
The amber indicator light
is on along with a two-tone
alarm and a red screen. To
clear the alarm, remove
power from the pump.
High Priority Alarm
The pump pauses or stops
if it is running. The pump
screen is red and the alarm
continues until a key is
pressed or the condition that
triggered it goes away.
Medium Priority Alarm
The pump does not stop if it
is running. The pump screen
is amber and the alarm
continues until a key is
pressed or the condition that
triggered it goes away.
Low Priority Alarm
The pump does not stop
if it is running. The pump
screen is blue and the alarm
continues for five seconds or
until a key is pressed or the
condition that triggered it
goes away.
Informational Message
The pump does not stop if
it is running. The message
appears in the status bar.
The alarm continues for
five seconds and may
be silent, requiring no
acknowledgement.