1.
Make sure the pump
is stopped and clamp
the tubing
2.
If the cassette is locked,
insert the pump key
and the turn the lock
counterclockwise into the
unlocked position
3.
Push down on the
cassette latch until the
cassette detaches
Ensure that the pump is stopped, the tubing is
disconnected from the patient, and the tubing
clamps are open.
After changing a cassette
1.
If a cassette is attached after the pump is powered
on, a Prime Tubing? screen will appear. Select Yes
(unlock the keypad if required)
2.
Select Prime
3.
Select Stop Priming when the air is removed or the
delivery will stop at 10mL (or 20mL if a high volume
set is attached)
No cassette is changed
1.
In the Tasks menu, press or to highlight Prime Tubing
and press Select (unlock the keypad if required)
2.
Select Prime
3.
Select Stop Priming when the air is removed or the
delivery will stop at 10mL (or 20mL if a high volume
set is attached)
Removing a Cassette
After attaching a new cassette
1.
The screen displays Reset reservoir volume to XX
mL? Select Yes to reset the volume or No to keep
the volume at the current setting
Without changing the cassette
1.
Stop the pump if it is running
2.
In the Tasks menu, press
or
to highlight Reset
Reservoir Volume and press Select
3.
The screen displays Reset reservoir volume to XX
mL? Select Yes to reset the volume
Resetting the Reservoir Volume
Priming the Tubing