background image

1. 

Make sure the pump  

is stopped and clamp  

the tubing

2. 

If the cassette is locked, 

insert the pump key  

and the turn the lock 

counterclockwise into the  

unlocked position

3. 

Push down on the  

cassette latch until the  

cassette detaches

Ensure that the pump is stopped, the tubing is 

disconnected from the patient, and the tubing  

clamps are open.

After changing a cassette

1. 

If a cassette is attached after the pump is powered 

on, a Prime Tubing? screen will appear. Select Yes 

(unlock the keypad if required) 

2. 

Select Prime

3. 

Select Stop Priming when the air is removed or the 

delivery will stop at 10mL (or 20mL if a high volume 

set is attached)

No cassette is changed

1. 

In the Tasks menu, press or to highlight Prime Tubing 

and press Select (unlock the keypad  if required)

2. 

Select Prime

3. 

Select Stop Priming when the air is removed or the 

delivery will stop at 10mL (or 20mL if a high volume 

set is attached)

 Removing a Cassette

After attaching a new cassette

1. 

The screen displays Reset reservoir volume to XX 

mL? Select Yes to reset the volume or No to keep  

the volume at the current setting

Without changing the cassette

1. 

Stop the pump if it is running

2. 

In the Tasks menu, press 

 or 

 to highlight Reset 

Reservoir Volume and press Select

3. 

The screen displays Reset reservoir volume to XX 

mL? Select Yes to reset the volume

Resetting the Reservoir Volume

Priming the Tubing

Summary of Contents for CADD

Page 1: ...B Display C Indicator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J J Cassette Keypad Lock K K Power Switch CADD Solis VIP Ambulatory Infusion Pump Quick Reference Card for Clinicians ...

Page 2: ...insert four AA batteries matching the and markings inside the battery compartment or insert a rechargeable battery pack 2 Close the compartment cover when the batteries are in place Insert Batteries 1 Press and hold the power switch 2 The pump carries out self tests and sounds six beeps when the tests are complete 3 Home screen is displayed Power On With the code 1 Press or until the first digit o...

Page 3: ...tings 6 Press continue to unlock the keypad using the security code or the pump key 7 Select the therapy press or to highlight the desired therapy and press Select 8 Select the qualifier press or to highlight the desired qualifier and press Select 9 Select the drug press or to highlight the desired drug and press Select 10 Confirm and review the settings press Yes if the information is correct and...

Page 4: ...back to the beginning 1 Clamp the tubing and open the cassette latch 2 Insert the cassette hooks into the hinge pins on the bottom of the pump Swing the cassette to the latch position 3 Place the pump upright on a firm flat surface and press down on the latch side of the pump so the cassette fits tightly against the pump 4 Lift the cassette latch into the closed position If you experience resistan...

Page 5: ...op at 10mL or 20mL if a high volume set is attached No cassette is changed 1 In the Tasks menu press or to highlight Prime Tubing and press Select unlock the keypad if required 2 Select Prime 3 Select Stop Priming when the air is removed or the delivery will stop at 10mL or 20mL if a high volume set is attached Removing a Cassette After attaching a new cassette 1 The screen displays Reset reservoi...

Page 6: ...patients to prime the tubing without having to enter a security code 1 In the Patient Permissions menu press or to highlight Priming Security On Off and press Select 2 Unlock the keypad 3 Press or to set the security to on security code required or off no security code required and select Save Delayed start security on off Setting this value to off allows patients to set delayed starts without hav...

Page 7: ...m continues until a key is pressed or the condition that triggered it goes away Low Priority Alarm The pump does not stop if it is running The pump screen is blue and the alarm continues for five seconds or until a key is pressed or the condition that triggered it goes away Informational Message The pump does not stop if it is running The message appears in the status bar The alarm continues for f...

Page 8: ...atteries or a rechargeable battery pack To start delivery good batteries must always be installed even when an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Select Acknowledge to clear the alarm or it will automatically clear after five seconds Recharge or change the rechargeable battery pack or replace the four AA batteries so...

Page 9: ...pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be due to a kink closed clamp or air bubble in the tubing between the fluid container and pump Delivery is paused and will resume if the occlusion is removed Remove the obstruction to resume operation The alarm will clear when the occlusion is removed You must acknowledge this alarm after it ...

Page 10: ...www smiths medical com customer support Smiths Medical is part of the global technology business Smiths Group plc Product s described may not be licensed or available for sale in all countries CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2016 Smiths Medica...

Reviews: