The delivery settings are patient-specifi c parameters 

of a therapy that are directly related to the drug being 

infused and can be edited within limits established in 

the protocol. 
To view and edit delivery settings:

1. 

Stop the pump if it is running 

2. 

In the Tasks menu, press 

 or 

 until View Delivery 

Settings is highlighted, then press Select 

3. 

Press 

 or 

 until the desired setting is highlighted, 

then press Select

4. 

If requested, unlock the keypad

5. 

Press 

 or 

 until the desired value appears on the 

screen, then select Save

Change any additional settings by scrolling through the 

remaining delivery settings and press Select to edit each 

setting as necessary.
Note: Editing individual delivery settings in Step or Taper 

mode will reset the infusion back to the beginning.

1. 

Clamp the tubing and open the cassette latch

2. 

Insert the cassette hooks into 

the hinge pins on the bottom 

of the pump. Swing the 

cassette to the latch position

3. 

Place the pump upright on a 

fi rm, fl at surface, and press 

down on the latch side of the 

pump so the cassette fi ts 

tightly against the pump

4. 

Lift the cassette latch into 

the closed position. If you 

experience resistance when 

lifting the cassette latch 

handle, do not force the latch. 

If the pump doesn’t latch 

easily, unlatch the cassette 

and repeat the process

5. 

Verify the cassette is attached properly. Looking from 

left to right, the top of the cassette should line up 

evenly with the bottom of the pump and be securely 

attached. If an uneven gap exists, unlatch the 

cassette and repeat the process

6. 

To lock the cassette, insert the pump key into the 

lock and turn it clockwise into the locked position

Editing Individual Delivery Settings

Attaching a Cassette

Summary of Contents for CADD

Page 1: ...B Display C Indicator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J J Cassette Keypad Lock K K Power Switch CADD Solis VIP Ambulatory Infusion Pump Quick Reference Card for Clinicians ...

Page 2: ...insert four AA batteries matching the and markings inside the battery compartment or insert a rechargeable battery pack 2 Close the compartment cover when the batteries are in place Insert Batteries 1 Press and hold the power switch 2 The pump carries out self tests and sounds six beeps when the tests are complete 3 Home screen is displayed Power On With the code 1 Press or until the first digit o...

Page 3: ...tings 6 Press continue to unlock the keypad using the security code or the pump key 7 Select the therapy press or to highlight the desired therapy and press Select 8 Select the qualifier press or to highlight the desired qualifier and press Select 9 Select the drug press or to highlight the desired drug and press Select 10 Confirm and review the settings press Yes if the information is correct and...

Page 4: ...back to the beginning 1 Clamp the tubing and open the cassette latch 2 Insert the cassette hooks into the hinge pins on the bottom of the pump Swing the cassette to the latch position 3 Place the pump upright on a firm flat surface and press down on the latch side of the pump so the cassette fits tightly against the pump 4 Lift the cassette latch into the closed position If you experience resistan...

Page 5: ...op at 10mL or 20mL if a high volume set is attached No cassette is changed 1 In the Tasks menu press or to highlight Prime Tubing and press Select unlock the keypad if required 2 Select Prime 3 Select Stop Priming when the air is removed or the delivery will stop at 10mL or 20mL if a high volume set is attached Removing a Cassette After attaching a new cassette 1 The screen displays Reset reservoi...

Page 6: ...patients to prime the tubing without having to enter a security code 1 In the Patient Permissions menu press or to highlight Priming Security On Off and press Select 2 Unlock the keypad 3 Press or to set the security to on security code required or off no security code required and select Save Delayed start security on off Setting this value to off allows patients to set delayed starts without hav...

Page 7: ...m continues until a key is pressed or the condition that triggered it goes away Low Priority Alarm The pump does not stop if it is running The pump screen is blue and the alarm continues for five seconds or until a key is pressed or the condition that triggered it goes away Informational Message The pump does not stop if it is running The message appears in the status bar The alarm continues for f...

Page 8: ...atteries or a rechargeable battery pack To start delivery good batteries must always be installed even when an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Select Acknowledge to clear the alarm or it will automatically clear after five seconds Recharge or change the rechargeable battery pack or replace the four AA batteries so...

Page 9: ...pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be due to a kink closed clamp or air bubble in the tubing between the fluid container and pump Delivery is paused and will resume if the occlusion is removed Remove the obstruction to resume operation The alarm will clear when the occlusion is removed You must acknowledge this alarm after it ...

Page 10: ...www smiths medical com customer support Smiths Medical is part of the global technology business Smiths Group plc Product s described may not be licensed or available for sale in all countries CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2016 Smiths Medica...

Reviews: