The delivery settings are patient-specifi c parameters
of a therapy that are directly related to the drug being
infused and can be edited within limits established in
the protocol.
To view and edit delivery settings:
1.
Stop the pump if it is running
2.
In the Tasks menu, press
or
until View Delivery
Settings is highlighted, then press Select
3.
Press
or
until the desired setting is highlighted,
then press Select
4.
If requested, unlock the keypad
5.
Press
or
until the desired value appears on the
screen, then select Save
Change any additional settings by scrolling through the
remaining delivery settings and press Select to edit each
setting as necessary.
Note: Editing individual delivery settings in Step or Taper
mode will reset the infusion back to the beginning.
1.
Clamp the tubing and open the cassette latch
2.
Insert the cassette hooks into
the hinge pins on the bottom
of the pump. Swing the
cassette to the latch position
3.
Place the pump upright on a
fi rm, fl at surface, and press
down on the latch side of the
pump so the cassette fi ts
tightly against the pump
4.
Lift the cassette latch into
the closed position. If you
experience resistance when
lifting the cassette latch
handle, do not force the latch.
If the pump doesn’t latch
easily, unlatch the cassette
and repeat the process
5.
Verify the cassette is attached properly. Looking from
left to right, the top of the cassette should line up
evenly with the bottom of the pump and be securely
attached. If an uneven gap exists, unlatch the
cassette and repeat the process
6.
To lock the cassette, insert the pump key into the
lock and turn it clockwise into the locked position
Editing Individual Delivery Settings
Attaching a Cassette