Reports can be viewed at any time, with the pump  

running or stopped.

1. 

In the Tasks menu, press 

 or 

 to highlight View 

Reports and press Select

2. 

Press 

 or 

 to highlight 

the desired report and  

press Select

3. 

Press Back to return to the 

Select Report menu and 

then Back again to return 

to the Tasks Menu

Reports

100 mL 

Stopped

Continuous 

10:28 AM       

8 Hour-Hydration

D5-0.45% NaCI 1000mL 

mL

Back 

Lock Keypad

Tasks

Press 'select' to begin

Display and Sound Settings

Change Time and Date

View Reports

View Advanced Tasks

Stop the pump if it is running. 

In the Advanced Tasks menu, 

press 

 or 

 to highlight 

Patient Permissions and  

press Select.

Priming security on/off

Setting this value to off allows 

patients to prime the tubing 

without having to enter a security code.

1. 

In the Patient Permissions menu, press 

 or 

 to 

highlight Priming Security On/Off and press Select

2. 

Unlock the keypad

3. 

Press 

 or 

 to set the security to on  

(security code required) or off (no security  

code required) and select Save

Delayed start security on/off

Setting this value to off allows patients to set delayed 

starts without having to enter a security code.

4. 

In the Patient Permissions menu, press 

 or 

  

to highlight Delayed Start Security On/Off and  

press Select

5. 

Unlock the keypad

6. 

Press 

 or 

 to highlight on (security code 

required) or off (no security code required)  

and select Save

Patient Permissions

100 mL 

Stopped

Continuous 

10:28 AM       

8 Hour-Hydration

D5-0.45% NaCI 1000mL 

mL

Back 

Lock Keypad

Advanced Tasks

Press 'select' to begin

Patient Permissions

Air and Occlusions Settings
Alarm Settings
Security Settings

Summary of Contents for CADD

Page 1: ...B Display C Indicator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J J Cassette Keypad Lock K K Power Switch CADD Solis VIP Ambulatory Infusion Pump Quick Reference Card for Clinicians ...

Page 2: ...insert four AA batteries matching the and markings inside the battery compartment or insert a rechargeable battery pack 2 Close the compartment cover when the batteries are in place Insert Batteries 1 Press and hold the power switch 2 The pump carries out self tests and sounds six beeps when the tests are complete 3 Home screen is displayed Power On With the code 1 Press or until the first digit o...

Page 3: ...tings 6 Press continue to unlock the keypad using the security code or the pump key 7 Select the therapy press or to highlight the desired therapy and press Select 8 Select the qualifier press or to highlight the desired qualifier and press Select 9 Select the drug press or to highlight the desired drug and press Select 10 Confirm and review the settings press Yes if the information is correct and...

Page 4: ...back to the beginning 1 Clamp the tubing and open the cassette latch 2 Insert the cassette hooks into the hinge pins on the bottom of the pump Swing the cassette to the latch position 3 Place the pump upright on a firm flat surface and press down on the latch side of the pump so the cassette fits tightly against the pump 4 Lift the cassette latch into the closed position If you experience resistan...

Page 5: ...op at 10mL or 20mL if a high volume set is attached No cassette is changed 1 In the Tasks menu press or to highlight Prime Tubing and press Select unlock the keypad if required 2 Select Prime 3 Select Stop Priming when the air is removed or the delivery will stop at 10mL or 20mL if a high volume set is attached Removing a Cassette After attaching a new cassette 1 The screen displays Reset reservoi...

Page 6: ...patients to prime the tubing without having to enter a security code 1 In the Patient Permissions menu press or to highlight Priming Security On Off and press Select 2 Unlock the keypad 3 Press or to set the security to on security code required or off no security code required and select Save Delayed start security on off Setting this value to off allows patients to set delayed starts without hav...

Page 7: ...m continues until a key is pressed or the condition that triggered it goes away Low Priority Alarm The pump does not stop if it is running The pump screen is blue and the alarm continues for five seconds or until a key is pressed or the condition that triggered it goes away Informational Message The pump does not stop if it is running The message appears in the status bar The alarm continues for f...

Page 8: ...atteries or a rechargeable battery pack To start delivery good batteries must always be installed even when an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Select Acknowledge to clear the alarm or it will automatically clear after five seconds Recharge or change the rechargeable battery pack or replace the four AA batteries so...

Page 9: ...pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be due to a kink closed clamp or air bubble in the tubing between the fluid container and pump Delivery is paused and will resume if the occlusion is removed Remove the obstruction to resume operation The alarm will clear when the occlusion is removed You must acknowledge this alarm after it ...

Page 10: ...www smiths medical com customer support Smiths Medical is part of the global technology business Smiths Group plc Product s described may not be licensed or available for sale in all countries CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2016 Smiths Medica...

Reviews: