background image

Screen is blank and alarm is sounding

Alarm Priority High. The pump was delivering and  

the batteries were removed or the battery door was 

opened. The pump has lost power and is no longer 

delivering. Clear the alarm by turning the pump back  

on, or the alarm will stop after the power has been off 

for a minimum of two minutes.

Air-in-line detected. Press “acknowledge” then 

prime tubing

Alarm Priority High. The air detector has detected air 

in the fluid path. The pump was delivering and is now 

stopped and will not run. Select Acknowledge to clear 

the alarm. If the fluid path contains air bubbles, close the 

clamps, disconnect the fluid path from the patient, then 

prime the tubing to remove the air and restart the pump. 

Battery depleted. Pump stopped.

Alarm Priority High. If the AC adapter is attached,  

select Acknowledge to clear the alarm. Remove 

the batteries and install four new AA batteries or a 

rechargeable battery pack. To start delivery, good 

batteries must always be installed even when an 

external source of power is connected. If appropriate, 

restart the pump.

Battery low. Replace battery.

Alarm Priority Low. Select Acknowledge to clear the 

alarm, or it will automatically clear after five seconds. 

Recharge or change the rechargeable battery pack or 

replace the four AA batteries soon. 

Downstream occlusion. Clear occlusion between 

pump and patient.

Alarm Priority High. The pump has detected high 

pressure, which may be resulting from a downstream 

blockage, kink in the fluid path, or a closed tubing clamp. 

Delivery pauses and resumes if the occlusion is removed. 

Remove the obstruction or select Stop Pump to silence 

the alarm for two minutes, then remove the obstruction 

and restart the pump. 

Reservoir volume low

Alarm Priority Medium or Low. The priority depends on 

how the alarm is programmed in Admin Settings. The 

level of fluid in the reservoir is low. Select Acknowledge 

to clear the alarm and prepare to install a new reservoir, 

if appropriate. 

Troubleshooting

Summary of Contents for CADD

Page 1: ...B Display C Indicator Lights D USB Port E Blue AC Power Light F AC Power Jack G Remote Dose Cord Jack H Keypad I Cassette Latch J J Cassette Keypad Lock K K Power Switch CADD Solis VIP Ambulatory Infusion Pump Quick Reference Card for Clinicians ...

Page 2: ...insert four AA batteries matching the and markings inside the battery compartment or insert a rechargeable battery pack 2 Close the compartment cover when the batteries are in place Insert Batteries 1 Press and hold the power switch 2 The pump carries out self tests and sounds six beeps when the tests are complete 3 Home screen is displayed Power On With the code 1 Press or until the first digit o...

Page 3: ...tings 6 Press continue to unlock the keypad using the security code or the pump key 7 Select the therapy press or to highlight the desired therapy and press Select 8 Select the qualifier press or to highlight the desired qualifier and press Select 9 Select the drug press or to highlight the desired drug and press Select 10 Confirm and review the settings press Yes if the information is correct and...

Page 4: ...back to the beginning 1 Clamp the tubing and open the cassette latch 2 Insert the cassette hooks into the hinge pins on the bottom of the pump Swing the cassette to the latch position 3 Place the pump upright on a firm flat surface and press down on the latch side of the pump so the cassette fits tightly against the pump 4 Lift the cassette latch into the closed position If you experience resistan...

Page 5: ...op at 10mL or 20mL if a high volume set is attached No cassette is changed 1 In the Tasks menu press or to highlight Prime Tubing and press Select unlock the keypad if required 2 Select Prime 3 Select Stop Priming when the air is removed or the delivery will stop at 10mL or 20mL if a high volume set is attached Removing a Cassette After attaching a new cassette 1 The screen displays Reset reservoi...

Page 6: ...patients to prime the tubing without having to enter a security code 1 In the Patient Permissions menu press or to highlight Priming Security On Off and press Select 2 Unlock the keypad 3 Press or to set the security to on security code required or off no security code required and select Save Delayed start security on off Setting this value to off allows patients to set delayed starts without hav...

Page 7: ...m continues until a key is pressed or the condition that triggered it goes away Low Priority Alarm The pump does not stop if it is running The pump screen is blue and the alarm continues for five seconds or until a key is pressed or the condition that triggered it goes away Informational Message The pump does not stop if it is running The message appears in the status bar The alarm continues for f...

Page 8: ...atteries or a rechargeable battery pack To start delivery good batteries must always be installed even when an external source of power is connected If appropriate restart the pump Battery low Replace battery Alarm Priority Low Select Acknowledge to clear the alarm or it will automatically clear after five seconds Recharge or change the rechargeable battery pack or replace the four AA batteries so...

Page 9: ...pump and reservoir Alarm Priority High Fluid is not flowing from the fluid container to the pump which may be due to a kink closed clamp or air bubble in the tubing between the fluid container and pump Delivery is paused and will resume if the occlusion is removed Remove the obstruction to resume operation The alarm will clear when the occlusion is removed You must acknowledge this alarm after it ...

Page 10: ...www smiths medical com customer support Smiths Medical is part of the global technology business Smiths Group plc Product s described may not be licensed or available for sale in all countries CADD and the Smiths Medical design mark are trademarks of Smiths Medical The symbol indicates the trademark is registered in the U S Patent and Trademark Office and certain other countries 2016 Smiths Medica...

Reviews: