36
EN
Power requirements
Device input voltage
21 VDC
Device input power
36 Watts
Mains adapter
Smith & Nephew REF 66801558
Input: 100–240VAC, 50/60HZ, 34–58VA
Output: 21VDC, 1.71A, 36W
Fuses
Internal electronic fuse, not user changeable
Power cord
Length: 1.8m maximum (Class II)
Dimensions
7 x 8.3 x 3.5in. / 175 x 210 x 85mm
Weight
2.4lbs / 1.1 kg
Battery
Operating time
20 hours (therapy)
Type
Lithium ion
Charging time
3 hours
Safety protection
Protection against electric shock
Class II
Patient protection
Type BF
Ingress protection
IP22
Environmental
Storage and transport
14 to 131°F / -10 to 55°C
Operational temperature
41 to 95°F / 5 to 35°C
Relative humidity
30% to 70% RH
Atmospheric pressure
700 mbar to 1,060 mbar
The time required for the device to warm
from 14°F / -10°C storage temperature
until the device can be used.
40 minutes at 68°F / 20°C
The time required for the device to cool
from 131°F / 55°C storage temperature
until the device can be used.
56 minutes at 68°F / 20°C
Compliance
Standards
CONFORMS TO
AAMI STD ES60601-1
IEC STD 60601-1
IEC STD 60601-1-8
IEC STD 60601-1-11
CERTIFIED TO CAN/CSA STD C22.2 No. 60601-1
Compliance
Standards
IEC STD 60601-1-2
10966 175239 Rev02 260917.indd 36
27/09/2017 09:09