35
EN
Specifications
Pressure
Setting
40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg:
tol/- 10 mmHg
Maximum
200mmHg
Alarms
General (all alarms)
Priority
Low
Auditory sound level
46dB
Over vacuum
Indicator color
Yellow
High flow/leak
Indicator color
Yellow
Alarm overall delay
60 seconds ± 5 seconds
Low vacuum
Indicator color
Yellow
Alarm overall delay
30 seconds ± 5 seconds
High vacuum
Indicator color
Yellow
Alarm overall delay
30 seconds ± 5 seconds
Blockage/canister full
Indicator color
Yellow
Alarm overall delay
120 seconds ± 5 seconds
Low battery
Indicator color
Yellow
Inactive
Indicator color
Yellow
Device not calibrated
Indicator color
Yellow
Device failed
Indicator color
Yellow
10966 175239 Rev02 260917.indd 35
27/09/2017 09:09