Battery power
Device can be used on battery power to provide the user greater mobility.
During therapy, if battery level is low, the device can be plugged into AC mains power without interrupting therapy.
Battery will recharge while therapy continues to be delivered.
Battery status indication
RENASYS™ GO contains a lithium ion rechargeable battery. There are several battery level indicators on this device:
Indicator
Battery level
Action
Device battery life will vary depending
on pressure setting selected. A fully
charged battery will typically deliver
therapy for approximately 21 to 36
hours if pressure setting is set between
40 to 120mmHg. At a pressure setting of
200mmHg, a fully charged battery will
typically deliver approximately 14 hours
of therapy.
Battery status top indicator will
illuminate solid green.
No action required.
Battery has up to
10 hours
active
therapy remaining.
Battery status top indicator will flash
green.
No action required.
Battery has up to
6 hours
active
therapy remaining.
Battery status top indicator will flash
green and battery status bottom
indicator will illuminate solid yellow.
No action required.
Battery has up to
3 hours
active
therapy remaining.
Battery status bottom indicator will
illuminate solid yellow.
Low battery alarm state (see alarm
section).
Audio alarm can be paused
by pressing the
Audio
pause
button on user
interface.
Plug device into AC mains
power to charge battery.
y
r
e
t
t
a
B
l
l
u
F
m
0
0
1
►
g
H
m
y
r
e
t
t
a
B
l
l
u
F
m
0
0
1
►
*
*
*
g
H
m
t
n
e
t
t
A
!
n
o
i
a
B
w
o
L
!
y
r
e
t
t
Display screen when device is
locked for homecare use.
Full charge and the Battery Full display may not be possible when charging at high ambient temperatures
(approximately 30°C/86°F). It is not necessary to continue charging for longer than 3 hours.
10966 175239 Rev02 260917.indd 17
27/09/2017 09:09