HRX-OM-Q026-B
Chapter 3 Transport and Setting Up
HRSH Series
3.5
Fill of circulating fluid
3-31
3.5.3
For option K [Fluid fill port]
Open the cap of the fluid fill port and fill the circulating fluid up to the level between the
“HIGH” and “LOW” of the fluid level gauge.
Fig. 3-22 Filling the fluid to the fluid fill port (An example)
z
Confirm that the fluid level is between “HIGH” and “LOW” level of the
fluid level gauge.
z
Be sure to connect the piping from the overflow port to the sump pit
z
Check the drain port is closed by the valve to prevent the supplied
circulating fluid from draining out.
Summary of Contents for HRSH150-A*-20 Series
Page 7: ...HRX OM Q026 B Contents HRSH Series Chapter 9 Product Warranty 9 1...
Page 8: ......
Page 177: ......