HRX-OM-Q026-B
Chapter 5 Display and setting of various functions
5.21
Communication function
HRSH Series
5-76
34.
Select contact input signal 1 delay timer of reading from the table below with [
▲
] key or
[
▼
] key, and confirm by pressing “SEL”.
Table 5.21-18
List of set value
Set value
Explanation
Initial value
(Default setting)
Setting and checking are not available unless
contact input signal 1 is external switch signal
input (N.O. type or N.C. type).
to
Setting of contact input signal 1 delay timer of
reading.
Set range is 0 to 300 sec.
Contact input signal 1 OFF detection timer
Setting and checking
35.
Press the [SEL] key once.
The set screen of the contact input signal 1 OFF detection timer is displayed on the
digital display.
PV
SV
36.
Select contact input signal 1 OFF detection timer from the table below with [
▲
] key or
[
▼
] key, and confirm by pressing “SEL”.
Table 5.21-19
List of set value
Set value
Explanation
Initial value
(Default setting)
Setting and checking are not available unless
contact input signal 1 is external switch signal
input (N.O. type or N.C. type).
to
Setting of contact input signal 1 OFF
detection timer
Set range is 0 to 10sec.
Contact input signal 2 Setting
and checking
37.
Press the [SEL] key once.
The set screen of contact input signal 2
is displayed on the digital display.
PV
SV
Summary of Contents for HRSH150-A*-20 Series
Page 7: ...HRX OM Q026 B Contents HRSH Series Chapter 9 Product Warranty 9 1...
Page 8: ......
Page 177: ......