2.4 SETUP AND CONNECTION
Before initially switching on the MCS I, leave the unit on the designated workspace for
at least 2 hours or more, to adjust to room temperature. Installing and initial adjusting
only may be operated by engineers specially trained by Slee. Before connecting the
instrument to the power supply, make sure the unit is adequate to your local electrical
power line. The unit is delivered “ready to start”. For the connection to your grounded
laboratory power supply use the delivered cable.
The MCS I can be used with a charcoal filter and has two powerful exhaust fans and so
it can be used even in laboratories without separate central exhaust-system. The filters
have to be exchanged at least once a year, depending on frequency of use. Fitting
replacement filters can be ordered at Slee. If you have an existing central exhaust-
system, you can connect the machine to this unit of course.
Leave space of at least 10cm from a wall or other units to all sides of
the MCS I, to guarantee an optimized air circulation.
For detailed information ask your local SLEE medical GmbH sales
representative.
Don’t put or store inflammable things to the direct neighbourhood of
the MCS I.
2.5 OPERATING CONDITIONS
•
The MCS I is concepted for use in combination with Slee´s
multistainer MAS, but may also be operated as stand-alone by
installing a separate loading station.
•
Operating of the MCS II is only intended within closed premises.
•
Field temperatures may differ from + 15°C to + 40°C (no big oscillations).
•
Max. allowed height over sea: 2000 m (EN 61010-1:2001).
•
Vibration free and weight-adequate work ground required.
•
Relative air humidity max. 80%, not condensing for temperatures up to 31°C,
Linearly decreasing to 50% relative humidity at 40°C.
•
No exposure to direct sunlight.
•
No variations of electrical power supply more than to 10%.