8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
Sirona Dental Systems GmbH
Operating Instructions CEREC AC
60
8.1.5.2.4
Refitting the prismatic tube
NOTE
Do not use CEREC 2 / CEREC 3 prismatic tubes.
NOTE
Risk of damaging the front lens or prism.
➢ The prismatic tube must not touch the front lens.
➢ Push the prismatic tube straight toward the camera body; do not
tilt it.
➢ Carefully refit the prismatic tube until it locks in place.
8.1.5.2.5
Prismatic tube (with sapphire glass) reprocessing overview
See detailed instructions after summary.
Prismatic tube for CEREC Bluecam with sapphire glass
Pre-Cleaning process
(while prismatic tube is attached to the camera)
1.
Clean with a cotton gauze moistened with tap water of potable water quality that meets Federal Clean Water
Standards and pH neutral soap
or
use a clean lint-free cloth which has been soaked in CaviCide
or
use Cavi Wipes
or
use pdCARE Wipes
or
use a clean lint-free cloth which has been soaked in > 60% isopropyl alcohol
2.
Dry the prismatic tube with a clean lint-free cloth.
HLD process
1.
Use CIDEX
®
OPA or Sporox II Sterilizing and
Disinfecting Solution as disinfectant for high-level
disinfection per manufacturer´s instructions.
2.
Leave the prismatic tube in CIDEX
®
OPA for 12
minutes or in Sporox II for 30 minutes at 20° C.
3.
Wipe off the prismatic tube. Use a cotton gauze
moistened with tap water of potable water quality
that meets Federal Clean Water Standards.
4.
Dry the prismatic tube with a clean lint-free cloth.
Dry heat sterilization process
➢ No sterilization possible.