background image

7500-700 Battery Charger

Instructions for Use

Summary of Contents for 7500-700

Page 1: ...7500 700 Battery Charger Instructions for Use...

Page 2: ...0 Battery Charger Instruction Manual Technical Description REF 7500 700 Ratings Class 1 Equipment Input Rating 100 240 V 50 60 Hz 0 30 Kw Output Rating 24 V DC 150 W Intended Use The MicroAire REF 750...

Page 3: ...liability can only be achieved when the equipment is connected to an equipment receptacle marked Hospital Only or Hospital Grade WARNING No modification of this equipment is allowed WARNING Prior to u...

Page 4: ...escription Use Standard Refer to Instruction Manual Booklet ISO 7010 M002 Indicates a MANDATORY action for the user to consult the Instructions For Use IFU Symbol must be blue as shown IEC 60601 1 200...

Page 5: ...bol is filled see Manufacturer symbol both the date of manufacture and the name address of the manufacturer may be combined in one symbol ISO 15223 1 20121 Manufacturer 3082 Indicates the medical devi...

Page 6: ...attery takes too long to begin or complete fast charge Remove and re insert the battery in a different slot to attempt charging again If three consecutive Charge Failed indications are received the ba...

Page 7: ...ve CAUTION Do not immerse in liquids to cool CAUTION Do not flash sterilize CAUTION Do not use Ethylene Oxide Sterilization CAUTION Do not use a washer sterilizer CAUTION Do not process in equipment u...

Page 8: ...s in service and or problems tracking your equipment outside the U S unless specified otherwise DISPOSAL RECYCLE Always follow the current local reccomendations and or regulations governing environmen...

Page 9: ...mits Compliance Mains terminal disturbance voltage conducted emissions CISPR 11 CISPR 11 Class A Complies with Class A Limits Electromagnetic radiation disturbance radiated emissions CISPR 11 CISPR 11...

Page 10: ...r 60 Hz Test Level 30 A m Complies Per Note Voltage Dips IEC 61000 4 11 Dip to 0V for 1 Cycle with 0 Degree Sync Angle Dip to 0V for 0 5 Cycles with Sync Angles of 0 45 90 135 180 225 270 315 degrees...

Page 11: ...IM 7500 700 Rev K 2021 05 11 of 12...

Page 12: ...05 2021 MicroAire Surgical Instruments LLC 3590 Grand Forks Boulevard Charlottesville Virginia 22911 USA Phone 800 722 0822 434 975 8000 Order Fax 800 648 4309 or 434 975 4131 www microaire com MediM...

Reviews: