IM-7500-700 Rev K 2021-05
10 of 12
EMC Test and Standard
Test Levels / Limits
Compliance
Immunity to conducted disturbances induced
by RF fields (conducted RF disturbance
immunity)- a .c . mains: IEC 61000-4-6
Input AC Power Cord
Frequency range: lS0kHz to 80MHz
Level: 3 V outside the ISM band, 6 Vin the ISM
bands between 0 .15 MHz and 80 MHz
Frequency step: 1%
Modulation: 80% Am at lkHz
Dwell Time: Minimum 1 s
Complies Per Note*
Power Frequency Magnetic Field Immunity:
IEC 61000-4-8
Frequency: 50 Hz or 60 Hz Test Level: 30 A/m
Complies Per Note*
Voltage Dips: IEC 61000-4-11
Dip to 0V for 1 Cycle with 0 Degree Sync Angle .
Dip to 0V for 0 .5 Cycles with Sync Angles of
0,45,90,135,180,225,270,315 degrees
Dip to 70% of 240V for 25 Cycles at 50Hz
Dip to 70% of 100V for 30 Cycles at 60Hz
Complies Per Note*
Complies Per Note*
Complies Per Note*
Complies Per Note*
Voltage Interruptions Immunity:
IEC 61000-4-11
Dip to 0V for Sms . (250 Cycles at (S0Hz) or 300
Cycles at 60 Hz)
Complies Per Note*
*NOTE: The 7500-700 Battery Charger has no Essential Performance as defined by ANSI/AAMI/ES60601-1 . Compliance
with Immunity Testing is defined as: at least one charger output will remain functional after the test .