IM-7500-700 Rev K   2021-05

6 of 12

BATTERY CHARGER SETUP AND OPERATION 

1 .  Plug in the Charger to a standard wall power supply .  

All LED lights will turn on for 10 seconds, off ten seconds, 

and on for ten seconds again, except for the Standby LED 

in the center . 

2 .  At the end of the LED test sequence, the single Green 

Standby LED in the center will turn on to indicate that 

the charger is ready . This light will remain on until the 

charger is unplugged from the power source .

OPERATION

1 .  Verify that the “Standby” LED is on .
2 .  Insert a MicroAire REF 6640-710 or REF 7505-710 Battery into any slot . The battery can only be inserted into the 

charger in one position .

3 .  If charging the REF 7500-620 Aseptic Battery, first place the REF 7500-625 Charger Adapter into the charger . Then 

place the REF 7500-620 into the REF 7500-625 .

4 .  The “Charging” LED corresponding to that slot will blink until the battery is ready for fast charge . If the battery is 

hot, or has a very low charge, the “Charging” LED may blink for some time . When fast charge begins the “ 

Charging” LED will be on continuously .

5 .  When the charging sequence is complete, the “Charge Complete” LED will illuminate .

6 .  The “Charge Failed” LED will illuminate if the battery takes too long to begin or complete fast charge . Remove and 

re-insert the battery in a different slot to attempt charging again . If three consecutive “Charge Failed”  

indications are received, the battery should be replaced .

WARNING: 

 

Prior to use, system components should be inspected and operated to detect any damage or  

 

 

malfunction . Do not use if damage is apparent .

NOTE:  

 

The Battery Charger is designed for use only with MicroAire Batteries REF 6640-710, REF 7505-710  

 

 

and REF 7500-620 Aseptic Battery using a REF 7500-625 charger adapter .

NOTE:  

 

All personnel should become familiar with the power equipment before it is set-up for use in any 

  

 

procedure . Personnel in-serviced should include, but not limited to, central processing personnel,  

 

 

members of the surgical team, and the bioengineering department .

Compatible Instruments

PRODUCT CATALOG NUMBER (REF) DESCRIPTION 

INSTRUCTIONS FOR USE WITH 

COMPLETE LIST OF ACCESSORIES

REF 6640-710

Small NiMH Battery Pack

IM-PWRBATT

REF 7505-710

Large NiMH Battery Pack

IM-PWRBATT

REF 7500-620

Aseptic Battery Pack

IM-ASEPBATT

REF 7500-625

Aseptic Battery Charger Adapter 

IM-ASEPBATT

Instruments and their accessories are sold separately .  Contact your MicroAire® Sales representative or MicroAire® Customer Service  

at 1-800-722-0822 for a complete list of accessories and instruments available . Outside the USA, contact your nearest MicroAire  

authorized distributor .

Summary of Contents for 7500-700

Page 1: ...7500 700 Battery Charger Instructions for Use...

Page 2: ...0 Battery Charger Instruction Manual Technical Description REF 7500 700 Ratings Class 1 Equipment Input Rating 100 240 V 50 60 Hz 0 30 Kw Output Rating 24 V DC 150 W Intended Use The MicroAire REF 750...

Page 3: ...liability can only be achieved when the equipment is connected to an equipment receptacle marked Hospital Only or Hospital Grade WARNING No modification of this equipment is allowed WARNING Prior to u...

Page 4: ...escription Use Standard Refer to Instruction Manual Booklet ISO 7010 M002 Indicates a MANDATORY action for the user to consult the Instructions For Use IFU Symbol must be blue as shown IEC 60601 1 200...

Page 5: ...bol is filled see Manufacturer symbol both the date of manufacture and the name address of the manufacturer may be combined in one symbol ISO 15223 1 20121 Manufacturer 3082 Indicates the medical devi...

Page 6: ...attery takes too long to begin or complete fast charge Remove and re insert the battery in a different slot to attempt charging again If three consecutive Charge Failed indications are received the ba...

Page 7: ...ve CAUTION Do not immerse in liquids to cool CAUTION Do not flash sterilize CAUTION Do not use Ethylene Oxide Sterilization CAUTION Do not use a washer sterilizer CAUTION Do not process in equipment u...

Page 8: ...s in service and or problems tracking your equipment outside the U S unless specified otherwise DISPOSAL RECYCLE Always follow the current local reccomendations and or regulations governing environmen...

Page 9: ...mits Compliance Mains terminal disturbance voltage conducted emissions CISPR 11 CISPR 11 Class A Complies with Class A Limits Electromagnetic radiation disturbance radiated emissions CISPR 11 CISPR 11...

Page 10: ...r 60 Hz Test Level 30 A m Complies Per Note Voltage Dips IEC 61000 4 11 Dip to 0V for 1 Cycle with 0 Degree Sync Angle Dip to 0V for 0 5 Cycles with Sync Angles of 0 45 90 135 180 225 270 315 degrees...

Page 11: ...IM 7500 700 Rev K 2021 05 11 of 12...

Page 12: ...05 2021 MicroAire Surgical Instruments LLC 3590 Grand Forks Boulevard Charlottesville Virginia 22911 USA Phone 800 722 0822 434 975 8000 Order Fax 800 648 4309 or 434 975 4131 www microaire com MediM...

Reviews: