IM-7500-700 Rev K 2021-05
6 of 12
BATTERY CHARGER SETUP AND OPERATION
1 . Plug in the Charger to a standard wall power supply .
All LED lights will turn on for 10 seconds, off ten seconds,
and on for ten seconds again, except for the Standby LED
in the center .
2 . At the end of the LED test sequence, the single Green
Standby LED in the center will turn on to indicate that
the charger is ready . This light will remain on until the
charger is unplugged from the power source .
OPERATION
1 . Verify that the “Standby” LED is on .
2 . Insert a MicroAire REF 6640-710 or REF 7505-710 Battery into any slot . The battery can only be inserted into the
charger in one position .
3 . If charging the REF 7500-620 Aseptic Battery, first place the REF 7500-625 Charger Adapter into the charger . Then
place the REF 7500-620 into the REF 7500-625 .
4 . The “Charging” LED corresponding to that slot will blink until the battery is ready for fast charge . If the battery is
hot, or has a very low charge, the “Charging” LED may blink for some time . When fast charge begins the “
Charging” LED will be on continuously .
5 . When the charging sequence is complete, the “Charge Complete” LED will illuminate .
6 . The “Charge Failed” LED will illuminate if the battery takes too long to begin or complete fast charge . Remove and
re-insert the battery in a different slot to attempt charging again . If three consecutive “Charge Failed”
indications are received, the battery should be replaced .
WARNING:
Prior to use, system components should be inspected and operated to detect any damage or
malfunction . Do not use if damage is apparent .
NOTE:
The Battery Charger is designed for use only with MicroAire Batteries REF 6640-710, REF 7505-710
and REF 7500-620 Aseptic Battery using a REF 7500-625 charger adapter .
NOTE:
All personnel should become familiar with the power equipment before it is set-up for use in any
procedure . Personnel in-serviced should include, but not limited to, central processing personnel,
members of the surgical team, and the bioengineering department .
Compatible Instruments
PRODUCT CATALOG NUMBER (REF) DESCRIPTION
INSTRUCTIONS FOR USE WITH
COMPLETE LIST OF ACCESSORIES
REF 6640-710
Small NiMH Battery Pack
IM-PWRBATT
REF 7505-710
Large NiMH Battery Pack
IM-PWRBATT
REF 7500-620
Aseptic Battery Pack
IM-ASEPBATT
REF 7500-625
Aseptic Battery Charger Adapter
IM-ASEPBATT
Instruments and their accessories are sold separately . Contact your MicroAire® Sales representative or MicroAire® Customer Service
at 1-800-722-0822 for a complete list of accessories and instruments available . Outside the USA, contact your nearest MicroAire
authorized distributor .