32
SIGMA Manual Number 41019
Revision AA
Preventive Maintenance Tests
CAUTION:
All testing and evaluations require fluid and air temperature of 72°F ±2°F.
NOTE:
The brand of IV sets being tested must be compatible with the pump under test. The
proper IV set calibration is listed on the label located on the top of the pump.
NOTE:
Sections identified with an asterisk (*) may be used as incoming test procedures.
Visual Inspection
1.
Inspect the pump for visual evidence of damage and/or defects to exterior components (i.e.
case, keypad and door).
■
The IV set tube channel must be free of foreign materials and dry for both testing and
use.
■
Any evidence of damage to, or repair of, these parts is reason to run the preventive
maintenance checks.
2.
Verify legibility of all labels.
3.
Replace labels as necessary.
Determine Software Version and Serial Number
When you press the
ON/OFF
key to turn the unit on, the Power Status screen displays the current
software version and serial number. You can also view pump and software information from the
User Option menu:
1.
From the Select Care Area screen, press
OPTIONS
.
2.
Select
User Options
, and press
OK
.
3.
Select
View Information
and press
OK
.
4.
Select
Pump Information
and press
OK
to display the Pump Info screen.
5.
Record the software version and serial number on the Preventive Maintenance Check
Sheet.
Keypad Test
The Spectrum keypad and display may be checked for proper operation as follows (The 1, 2, 3, and
5 keys will be verified during subsequent testing):
1.
Press the
ON/OFF
key to turn the unit on. Select any
Care Area
and Press
OK
.
2.
Select any
Drug Name
and Press
OK.
3.
Press
OK
to select
Primary Bag
and Press
OK
to select mL/hr, and then enter “4” and “6”
using the keypad and observe the
Rate mL/hr
field to be “46”.
4.
Press the
CLEAR
soft key and verify that the
Rate mL/hr
field is cleared to 0.
5.
Enter “7”, “8”, and “9” using the keypad and observe the
Rate mL/hr
field to be “789”.