23
Wireless Networking
SIGMA Manual Number 41019
Revision AA
WEPKEYINDEX=0
# WEPKEY may be blank or 10 (64-bit) or 26 (128-bit) hex (0-1 and a-f)
characters long
WEPKEY=
# WPAENCRYPTION=0 Any
# WPAENCRYPTION=1 WEP
# WPAENCRYPTION=2 TKIP
# WPAENCRYPTION=3 CCMP (AES)
# WPAENCRYPTION=4 Open (no encryption)
WPAENCRYPTION=0
# WPAPSK must be blank if WPA PSK is not used
# WPAPSK may 64 hex (0-1 and a-f) characters long to specify a PSK
# WPAPSK may be 8-63 ascii characters long to specify a passphrase
WPAPSK=
# EAP Authentication 32-bit bitmask (decimal format, hex for reference
only)
# 0 (0x00000000) - None
# 1 (0x00000001) - LEAP
# 2 (0x00000002) - PEAP
# 4 (0x00000004) - TLS
# 8 (0x00000008) - TTLS
# 256 (0x00000100) - GTC
# 512 (0x00000200) - MD5
# 1024 (0x00000400) - MSCHAPv2
# 2048 (0x00000800) - OTP
# 65536 (0x00010000) - CHAP
# 131072 (0x00020000) - MSCHAP
# 262144 (0x00040000) - MSCHAPv2
# 524288 (0x00080000) - PAP
AUTHENTICATION=0
# IDENTITY= 802.1x Identity (username)
IDENTITY=
# PASSWORD= 802.1x Password
PASSWORD=
Network Settings
This option shows the basic network settings for the pump, and also allows the pump’s network con-
nection to be enabled or disabled. The following fields are available:
■
Networking – allows the pump’s network connection to be turned on or off. This setting is
stored in the pump after the pump is turned off, although the pump’s network connection is
always turned off when the pump power is turned off, including when the pump is turned
off but in battery charge mode.
■
SIGMA GW – (read-only, set via network configuration file) the IP address or DNS name of
the network host, also known as the Sigma Gateway server. Either this item or ‘Multicast’
may be set, but not both.
■
Multicast – (read-only, set via network configuration file) the IP address of the multicast
group that the pump joins to learn the location of the host. Either this item or ‘SIGMA GW’
may be set, but not both.
■
Device ID – (read-only, set via network configuration file) the facility-defined device identi-
fier for this pump. This value should be unique for each pump on a facility’s network.