SIGMA Manual Number 41019
Revision AA
LOGGING
The SIGMA Model Spectrum Infusion pump provides three logs that can be used when trouble-
shooting issues with the pump:
■
History Log – contains the entire history log for the pump.
■
System Error Log – contains only the system errors recorded in the history log.
■
Drug Error Log – contains only the events associated with programming an infusion;
including any drug limits that may have been exceeded during setup.
You can view these logs on the pump screen from the User Options menu (see the User options sec-
tion of the SIGMA Spectrum Operator’s Manual).
Downloading the History Log
In addition to viewing the logs on the pump screen, you can send the entire history log out of the
pump via the Infrared Data Association (IrDA) port using the Dump History Log option.
NOTE:
This procedure applies to the transfer of the history (event) log contents from a SIGMA
Spectrum Infusion pump with V5.00.06 Operating Software to a PC via the IrDA port.
Equipment List
■
Personal computer (PC) running Windows XP or Windows 2000 with enabled IrDA com-
munications
■
Optional equipment list (if IrDA-equipped PC is not available)
■
PC running Windows XP or Windows 2000 with SD Card Reader or,
■
USB external SD card reader/writer, Ezonics™ or equivalent
■
SD card, Kingston™ or equivalent, minimum size 512MB
■
PN 55098 – PDA, PalmOne™ Tungsten™ E2 or equivalent with a file transfer utility
installed (McFile version 2.2sigma, or equivalent) for transferring data
Procedure
To download the history log:
1.
Access the History Log option on the pump:
a. If the pump is running, press
STOP
, press
HELP
and press
OPTIONS
.
b. Select the
User Options
menu item.
c. From the User Options menu, select the
View Information
.
d. Select
History Log
.
e. Select
Dump History Log
.
■
If you are transerring the log directly to a PC
, proceed to step 2.
■
If you are transerring the log to a PDA then to a PC
, proceed to step 3.