4
SIGMA Manual Number 41019
Revision AA
CAUTION:
-
Federal (USA) law restricts this device to sale by or on the order of a
physician or other licensed practitioner.
-
Operate only on the AC line voltage for which the AC power adaptor is
marked.
-
The AC power adaptor should be plugged in during storage to
maintain proper battery charge level.
-
Disconnect from AC power source and remove the battery pack prior
to servicing this pump.
-
Routine cleaning and periodic maintenance is needed to assure that the
pump remains safe and functional. Refer to “Service and
Maintenance” on page 26, and “Preventive Maintenance” on page 31.
-
Clean up solution spills as soon as possible using a damp cloth or
sponge. Dry thoroughly before use. Keep the upstream and
downstream sensors’ tube channels clean and dry at all times during
pump operation. Do not use hard or sharp objects to clean sensor tube
channel areas. Failure to comply with this requirement may cause the
pump to fail. Refer to “Service and Maintenance” on page 26.
-
Static sensitive electronics used in SIGMA infusion pumps may be
damaged by electrostatic discharge. Service technicians must follow
proper ESD procedures when working on pumps. Refer to MIL-STD-
1686 for protective handling procedures.
-
Do not use a pen or any other sharp object to actuate the keyboard
switches. Replace torn or punctured membranes immediately, to assure
solutions won't internally damage the membrane switch and prevent
proper operation of the SIGMA Spectrum. (Replacement should only
be performed by authorized personnel).
-
Use only filter sets that are specifically listed in the Operator's Manual
for use with Spectrum pumps.
NOTE:
For a list of serviceable parts, contact SIGMA.