34
SIGMA Manual Number 41019
Revision AA
4.
Verify that
DS Occlusion Auto Restart
is set for “on” and verify
US/Air Alarm Confirm
is
set for “on”.
5.
Press the
EXIT
soft key two times to return to the
OPTIONS
screen.
6.
On the options screen, select
User Options
and press
OK.
7.
On the User Options screen, select
Alarm Settings
and press
OK
.
8.
Set the
DS Pressure Limit
to “med”.
9.
Press the
EXIT
soft key three times to leave the Alarm Settings, User Options, and Options
screens.
10. Select a
Care Area
and press
OK
.
11. Press
OK
again and the desired drug list will appear, select
BASIC
and then press
OK
.
12. Select
Primary Bag
and press
OK
.
13. For
Mode
select
mL/hr
and press
OK
.
14. Set the
Rate
to
100 mL/hr
and set the
VTBI mL
to
50 mL
.
15. (Skip this step if IV set is already loaded in pump) Load the IV set in accordance with the
directions on the
LOAD SET
screen on the pump.
16. Connect a pressure gauge to the end of a fluid-filled new IV set with a length of ≈48” of
tubing between the pump and gauge.
17. Press the
RUN/STOP
key to start the pump.
NOTE:
Pumps with v6.00.00 and later software versions will display the Check
Flow screen once the pump is started. For purposes of this test, press
YES
to proceed to the normal Run screen.
18. The pump may indicate BAG NEAR EMPTY <30 MIN REMAIN. Press
OK
to silence this
alarm.
19. Verify the pump displays a
DOWNSTREAM OCCLUSION
alarm at a pressure within the
13-psi ±6 occlusion trip pressure setting.
20. Remove the IV set from the gauge and ensure the pump Auto Restarts.
21. Press the
RUN/STOP
key to stop the pump and Remove pressure gauge and re-set the test
IV set to continue to the next test.
Air Detection Test*
If continuing this test from the Down Stream (Distal) Occlusion Sensor Test Start with Step 9, if you
wish to conduct a standalone test start with step 1.
NOTE:
Confirm test failures with a new unused IV set.
1.
Press the
ON/OFF
key to turn the unit on. Select any
Care Area
and Press
OK.
2.
From the Drug List, select
BASIC
and then press
OK
.
3.
Select Primary Bag and press
OK
.