38
SIGMA Manual Number 41019
Revision AA
Pass criteria for a used battery
≥6 hours from “Start to dead” and ≥31 minutes from “Low to dead”.
NOTE:
If a wireless battery module is installed, the pump used battery pass criteria is ≥3 hours
from “Start to dead” and ≥31 minutes from “Low to dead”.
Pass criteria after installation of a new battery
≥8 hours from “Start to dead” and ≥31 minutes from “Low to dead”.
NOTE:
If a wireless battery module is installed, the pump new battery pass criteria is ≥4 hours
from “Start to dead” and ≥31 minutes from “Low to dead”.
The battery capacity is less than the pass criteria, the battery is reaching end of life. See the Battery
Removal and Replacement section of this manual for replacement. If a wireless battery module is
installed, see “Wireless Module Battery Cell Removal and Replacement” on page 28.