Ionics Instruments
2004
Page 4-4
DLM 30007-08 Rev. A
Plumbing
Installing On-Line Sample Inlet System –
1. Install a Swagelok adapter in the sampling port to permit attachment
of 1/4"-OD Teflon tubing and connect the tubing (end without filter)
to the adapter.
2. Connect the 1/4" Teflon tubing with the in-line filter to the inlet of the
pressure regulator.
WARNING
Operation of the analyzer without the in-line filter on the sample inlet line will
damage the analyzer and void the warranty. To avoid damaging the analyzer,
install the filter and replace as needed.
WARNING
Do not adjust the pressure regulator on the sample inlet system. The pressure
has been preset to 5 psi and changing the inlet pressure can damage the analyzer
and void the warranty.
WARNING
Do not run the analyzer with the sample inlet line capped off as this can result
in loss of deionized water.