Ionics Instruments
2004
Page 8-10
DLM 30007-08 Rev. A
Pull Nut
Back to Plug
Pull Up on Plug and
Nut to Clear the DI
Reservoir
DI
Reservoir
Mixed Bed Resin
Lift Plug and Nut to
Clear Hinge
FIGURE 8-3: Lamp Removal
6. Pull nut from lamp body up onto lamp cord all the way back to the plug.
This will help in removing the lamp from the enclosure.
7. Carefully pull the lamp out of the quartz coil and enclosure, being careful
not to break the quartz coil. Pull the nut and plug past the top of the DI
reservoir first, then carefully manipulate the lamp over the top of the
reservoir. Lifting on the plug will help position the lamp so that it clears the
top of the reservoir. Gently twisting the lamp back and forth will help to
remove the lamp from the enclosure.
8. Pull the plug and nut away from the hinge on the top cover so that the cord
and lamp will clear the hinge.
9. Put on the gloves supplied in the box. Remove the new lamp from the
packet. Remove the nut and ferrules from the zip-top bag and slide them
up the lamp housing until they are flush with the heat shrink tubing. Be
careful when inserting the lamp into the instrument as the nut and ferrules