Ionics Instruments
2004
A-3
DLM 30007-08 Rev. A
MAIN menu, the SELECT FREQUENCY CALIBRATION CONSTANTS menu
is displayed.
50 HZ CALIBRATION
ENTER
CLEAR
60 HZ CALIBRATION
FIGURE A-3: Select Frequency Menu for Sievers 810 TOC Analyzer
Use the arrows ( or ) to select the calibration constants for the operating
frequency and press ENTER.
The calibration constants must be selected
for the operating frequency.
The CALIBRATION menu is then displayed.
The procedures for calibrating the instrument are the same as described in the
Calibration chapter. As a reminder that the calibration constants are
dependent on the AC frequency, the CALIBRATION CONSTANT menu will
display the frequency as shown in Figure A-4 for the TC Calibration Constant.
TC CALIBRATION CONST
1.19810000
ENTER
CLEAR
ENT = SAVE CHAR/NUM
CLR = RESTART/EXIT
ARROWS = CHANGE CHAR
50 HZ
FIGURE A-4: TC Calibration Constant Menu for Sievers 810 TOC Analyzer
All other menus and operating procedures for the 810 and 810AS models are
identical to the 800 and 800AS TOC analyzer.