10. Reconnect the 1/4" Teflon tubing from the sampling port adapter to
the inlet of the in-line filter. The in-line filter has arrows on the body of
the filter that indicate the direction of flow through the filter. Tighten
the nut 1/4 turn past finger tight.
11. Reconnect the 1/4" Teflon tubing to the outlet of the filter and tighten
the nut 1/4 turn past finger tight.
12. Turn on the water supply at the sampling port prior to starting the
analyzer.
Replacing the Reagents
The oxidizer and acid reagents
must
be purchased from Ionics Instruments to
ensure product purity. The use of reagents from other sources or failure to
replace the reagents on the prescribed replacement schedule will invalidate
the warranty.
WARNING
Reagent containers are for single-use only – do not refill. Refilling or reusing
reagent containers will void all instrument and parts warranties and nullify any
performance claims.
WARNING
Installation of the reagents requires access to the inside of the analyzer. To
avoid potentially dangerous shock, disconnect the power cord before opening
the analyzer's top panel.
WARNING
To avoid exposure to the chemical reagents, wear acid-resistant gloves,
protective clothing, and safety goggles or a face shield when changing the
reagent supplies.
To replace the reagents:
1. Turn the main power switch off and unplug the power cord.
Summary of Contents for TOC 800
Page 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Page 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Page 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Page 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Page 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...