Ionics Instruments
2004
Page 10-8
DLM 30007-08 Rev. A
If the fatal error involves the servo mechanism for the syringe pumps,
remove the AC power cord, open the top of the analyzer and check
for obstructions in the mechanism. Check to make sure that the
ribbon cable, located on the left-hand side of the syringe, is plugged
into the socket. Remove any obstruction, reconnect the cable, and
restart the analyzer.
Warnings do not stop the analyzer but will cause the warning number to be
displayed on the RUNTIME screen.
TOC =
3.67 PPB
05:00
ENTER
CLEAR
DEIONIZED WATER
W12 W13
FIGURE 10-2: TOC Measurement Screen with Error Messages
The numbers for the last two warnings are displayed in the upper left-hand
corner of the screen. In this example, the acid and oxidant reagents need to
be replaced soon. To get information on the warnings and corrective actions:
1. Press ENTER to switch to the RUNTIME menu.
2. Scroll to ERRORS and press ENTER.
3. Scroll to the desired error number and press ENTER.
4. Perform any actions suggested in the Error Information menu.
If the UV lamp is turned off, the error message "–UV" will be displayed in the
upper left-hand corner of the screen. This warning is not written to the error
stack and is only a reminder if the lamp has not been turned on after
calibration of the TC zero offset.
Summary of Contents for TOC 800
Page 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Page 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Page 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Page 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Page 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...