f. Locate the oxidizer supply (the container on the right-hand
side) and remove it from the enclosure by lifting up on the
front of the container (the end opposite the Mininert valve),
pulling the container out of the enclosure while sliding the
valve out from under the tubing and stainless steel cross.
g. Close the Mininert valve on the container by pressing on
the red button to slide the stainless steel bar fully inward.
WARNING
Oxidizer may spill from the container if you do not close the valve.
h. Disconnect the oxidizer container from the inlet line by
grasping the container and the brown nut and unscrewing
the container counterclockwise while holding the brown
knurled nut. Turning the container rather than the nut will
prevent the tubing from twisting.
i. Store the reagent containers in a refrigerator.
j. Place a paper towel or Kimwipe around the acid and
oxidant inlet line to absorb any liquid.
k. Replace the top cover on the reagent enclosure and secure
with the two thumb-screws located at the back of the
reagent enclosure. Close the analyzer top cover panel.
The analyzer is now ready for repackaging. To prevent the deionized water
loop from freezing during shipment, two Warm Packs are included in the
accessories kit that came with the analyzer.
Warm Packs must be used if
returning the analyzer between September and May
. If Warm Packs are
not available, please contact Ionics Instruments for replacement.
Summary of Contents for TOC 800
Page 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Page 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Page 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Page 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Page 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...