12. Disconnect the oxidizer container from the inlet line by grasping the
container and the brown nut and unscrewing the nut
counterclockwise while holding the container. Turning the nut rather
than the container will prevent the tubing from twisting.
13. Use DI water and a paper towel to remove any liquid or solids from
the nut/ferrule and connect the new OXIDIZER reagent container to
the OXIDIZER inlet line by carefully screwing the container and
Mininert valve clockwise onto the nut and tighten fingertight. Do not
overtighten the nut.
14. Open the Mininert valve on the oxidizer container.
15. Install the oxidizer container on the right side of the enclosure with
the valve facing the rear of the analyzer.
NOTE
The used reagent containers may still have a small volume of 6 M H3PO4
(ACID) or 15% (NH4)2S2O8 (OXIDIZER) inside and should be disposed of
properly. Consult state and local regulations.
16. Replace the reagent enclosure top cover and secure with the thumb
screws.
17. Record the installation date for the reagents in the service log.
If you need to replace the UV lamp, follow the procedures listed in the next
section. Otherwise, close the top panel and restart the analyzer to enter
the installation date for the reagents in the CONSUMABLES menu.
18. Plug in the power cord and turn the power switch on. The analyzer
will start initialization.
19. Stop the analyzer by selecting STOP TOC.
20. Follow the procedures outlined later in this chapter to enter the
installation date for the reagents in the CONSUMABLES menu.
Summary of Contents for TOC 800
Page 6: ...Ionics Instruments 2004 Page v DLM 30007 08 Rev A Declaration of Conformity ...
Page 22: ...Ionics Instruments 2004 Page 1 1 DLM 30007 08 Rev A ...
Page 209: ...Ionics Instruments 2004 Page 10 1 DLM 30007 08 Rev A ...
Page 218: ...Ionics Instruments 2004 Page 11 9 DLM 30007 08 Rev A ...
Page 252: ...Ionics Instruments 2004 Page E 8 DLM 30007 08 Rev A ...