Ionics Instruments
2004
Page 5-12
DLM 30007-08 Rev. A
select to display TOC, IC, or TC (the factory default is TOC). All three
measurements are stored on the RAM card and available in HISTORY.
The DISPLAY option affects the Measurement screen only. To change the
display output:
1. Use the arrows ( or ) to scroll to DISPLAY. Press ENTER. The
Set Display Mode screen is displayed:
SET DISPLAY MODE
ENTER
CLEAR
CURRENT: TOC DISPLAY
NEW: IC DISPLAY
ENT = CHANGE CLR = EXIT
ARROWS = SELECT TYPE
FIGURE 5-12: SET DISPLAY MODE Screen
2. Use the arrows ( or ) to select the measurement to be displayed
(TOC, IC or TC). Press ENTER. The Current display will change to
indicate the new display option and the program will prompt for
confirmation. Press ENTER to return to the OUTPUTS menu.
3. Press CLEAR twice to return to the SETUP menu.
Warning 42
The accuracy of the TOC measurement is reduced when the TOC/IC ratio is
≤
0.1; for improved accuracy, use the IC Removal Module to reduce the
concentration of IC in the sample. After each measurement, the analyzer will
calculate the TOC/IC ratio and if this value is less than or equal to 0.1, a