Technical data
11.2 Digital measuring error
SITRANS TH200/TH300
78
Operating Instructions, 06/2010, A5E00393069-03
11.2
Digital measuring error
Digital measuring errors for resistance thermometers
Input point
Measuring range in °C (°F)
Minimum measuring
span in °C (°F)
Digital accuracy in °C
(°F)
According to IEC
60751
Pt25
-200 ... +850 (-328 ... +1562)
10 (18)
0,3 (0.54)
Pt50
-200 ... +850 (-328 ... +1562)
10 (18)
0,15 (0.27)
Pt100 ... Pt200
-200 ... +850 (-328 ... +1562)
10 (18)
0,1 (0.18)
Pt500
-200 ... +850 (-328 ... +1562)
10 (18)
0,15 (0.27)
Pt1000
-200 ... +350 (-328 ... +662)
10 (18)
0,15 (0.27)
According to
JIS C1604-81
10 (18)
Pt25
-200 ... +649 (-328 ... +1200)
10 (18)
0,3 (0.54)
Pt50
-200 ... +649 (-328 ... +1200)
10 (18)
0,15 (0.27)
Pt100 ... Pt200
-200 ... +649 (-328 ... +1200)
10 (18)
0,1 (0.18)
Pt500
-200 ... +649 (-328 ... +1200)
10 (18)
0,15 (0.27)
Pt1000
-200 ... +350 (-328 ... +662)
10 (18)
0,15 (0.27)
Ni25 ... Ni1000
-60 ... +250 (-76 ... +482)
10 (18)
0,1 (0.18)
Digital measuring errors for resistance-type transmitters
Input point
Measuring range
[Ω]
Minimum span in [Ω] Digital accuracy [Ω]
Resistance
0 ... 390
5
0,05
Resistance
0 ... 2200
25
0,25
Summary of Contents for SITRANS T4200
Page 7: ...Table of contents SITRANS TH200 TH300 6 Operating Instructions 06 2010 A5E00393069 03 ...
Page 71: ...Service and maintenance SITRANS TH200 TH300 70 Operating Instructions 06 2010 A5E00393069 03 ...