Explosion protection
5.2 FM
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
85
Connection cable
Table 5- 7
Electric connections for connection cable analog and Modbus, valid for FM product
versions
Terminals in receiver junction board
Function/Power
Ethernet cable
1
+
2
-
Supply voltage
19 ... 30.2 V, 10 VA
1)
3
4
Closed when energized
4)
Digital OUT 0 (relay)
30 V, 0.5 A
3)
5
6
Closed when energized
4)
Digital OUT 1 (relay)
30 V, 0.5 A
3)
7
+
8
-
Digital IN 0
0 ... 30 V
2)
9
+
10
-
Analog OUT 0 (Measurement)
30 V, 24 mA
3)
11
+
12
-
Analog OUT 1 (Measurement)
30 V, 24 mA
3)
13
PROFIBUS A-Line
(RxD/TxD_N - data
inverted)
Modbus D1
(RxD/TxD_N - data
inverted))
14
PROFIBUS B-Line
(RxD/TxD_P - data
not inverted)
Modbus D0
(RxD/TxD_P - data
not inverted)
15
PROFIBUS / Modbus shield
RS-485
(PROFIBUS
5)
/ Modbus)
- 7 ...+ 12 V DC
16
Tx+
White/Orange
17
Tx-
Orange
18
Rx+
White/Green
19
Rx-
Ethernet
6)
Green
20
+
21
-
Analog IN 0 (Temperature)
0 .. 30 mA
2)
22
+
23
-
Analog IN 1 (Pressure)
0 ... 30 mA
2)
24
Ground
25
Ground
Chassis
Ground
Gland
Ground
Shield
1)
Maximum possible power consumption.
2)
Maximum input values.
3)
Maximum output values. The following figures show how the outputs are used.
4)
Supported relay modes: "Normally Energized" (default), "Normally De-Energized". For configuration, see
section Configuration (Page 116).
5)
Certified PROFIBUS wiring recommended, see section PROFIBUS DP interface (Page 129)
6)
We recommend not to connect the connection cable directly to Ethernet connectors. Instead the sensor
connection cable kit should be used for this purpose.