Operation
7.3 Menu system
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
117
Path length
View/edit path length in meters with three decimal digits.
Temperature correction
This menu is dynamic; at first level you're asked to "Set source" which then displays a list of
all available temperature inputs.
The example in the following display sequence shows the steps to change the source from
analog input to manual input.
Parameter "Temperature correction" for
analog input
Parameter "Set source" in a PROFIBUS device
configuration
Parameter "Set source" in a Modbus
device configuration
Paramater "Temperature correction" for manual
input
Pressure correction
Settings, sequences, and parameters are similar to those of the temperature correction
menu.
Integration time
This menu is dynamic and depends on the application. Integration time settings are possible
for the following components:
●
Measurement component (e.g. O
2,
CO)
●
Absolute transmission
●
Temperature
●
Pressure
An averaging filter is connected to each component. For each component the filter
parameter "integration time" can be used to balance the needs of response time and noise
reduction.