Alarm, error, and system messages
10.2 Alarm list
SITRANS SL
174
Operating Instructions, 12/2010, A5E01132948-04
Symbol/Group
Error
Code
Alarm (description)
Explanation
/Action
E57
EEPROM configuration missing
EEPROM configuration missing.
Contact Siemens support.
E58
Save settings failed
Save settings failed.
Contact Siemens support.
E09
Save settings active
Configuration changed
E55
Analog output test mode
The test mode for analog outputs is active
E56
Digital output test mode
The test mode for digital outputs is active
E25
PROFIBUS configuration error
PROFIBUS configuration fault.
Contact PLC support.
E26
PROFIBUS connection error
PROFIBUS Connection Error.
Check that PROFIBUS cables are properly connected.
E27
PROFIBUS master switch off
PROFIBUS Master Switch Off.
Contact PLC support. Check the PROFIBUS master
switch.
E28
PROFIBUS internal error
PROFIBUS internal error.
Contact Siemens support.
E61
LUI unlocked
The LUI is unlocked and has no password protection.
Lock the LUI.
E60
Watchdog reset triggered
The analyzer has been reset due to a triggered watchdog
event.
This alarm is automatically cleared at the screen after
restart of the system. The alarm will still be visible in the
logbook.