Remote operation
7.1 Operating via SIMATIC PDM
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
71
View saved echo profiles
To view a saved profile, click on the tab View Saved Echo Profiles.
Echo profile data logging
You can store up to 60 profiles at a selected interval (maximum 60 minutes). Inside Echo Profile
Utilities, in the Echo Profile Time Based Storage window:
●
Enter the desired interval between stored profiles.
●
Enter the maximum number of profiles to be stored (maximum 60).
●
Click on Start. A message appears warning of the time delay and warning that all previous
saved profiles will be overwritten. Click on OK to proceed. The new profiles will be saved with
their date and time.
●
Click on the tab View Saved Echo Profiles to view the stored profiles.