Remote operation
7.1 Operating via SIMATIC PDM
SITRANS LR250 (PROFIBUS PA)
Operating Instructions, 01/2014, A5E32221386-AB
63
Quick start
Note
•
A reset to Factory Defaults should be performed before running the Quick Start Wizard if
device has been used in a previous application. See Master Reset via SIMATIC PDM
(Page 82).
•
The Quick Start wizard settings are inter-related and changes apply only after you click
on FINISH AND DOWNLOAD at the end of the last step to save settings offline and
transfer them to the device.
•
Do not use the Quick Start Wizard to modify individual parameters: for quick access to
echo profile parameters, see Echo Profile via SIMATIC PDM (Page 70) or see Parameter
Reference (Page 93) for a complete list. (Perform customization only after the Quick Start
has been completed.)
•
Click on BACK to return and revise settings or CANCEL to exit the Quick Start.
•
For a vessel with obstructions see Auto False Echo Suppression via SIMATIC PDM
(Page 73).
Launch SIMATIC PDM, open the menu Device – Wizard - Quick Start, and follow steps 1 to 5.