Installing/mounting
4.4 Flange bolting, Flanged encapsulated antenna only
SITRANS LR250 (PROFIBUS PA)
24
Operating Instructions, 01/2014, A5E32221386-AB
Recommendations for flange bolting:
●
Use cross-pattern sequence as shown
●
Check uniformity of the flange gap
●
Apply adjustments by selective tightening if required
●
Torque incrementally until desired value is reached
●
Check/re-torque after 4 to 6 hours
●
Check bolts periodically, re-torque as required
●
Use new lens, O-ring and spring washers after removal from installation.