Modbus
Address
Parameter
Default
value
Value
range
Access
level
Description
40265
216
Path 2 fault hours
e.g. 36 h
Read;
Read
Total hours fault active.
Valid only for 2-path solutions and if
alarm is enabled
40266
217
Path 2 fault counter
e.g. 4
0 to 65535 Read
Total number of faults.
Valid only for 2-path solutions and if
alarm is enabled
40267
218
Path 2 fault appears
Last fault
appear
date
dd-mm-
yyyy,
hh:mm:ss
Read
First time the fault appeared.
Valid only for 2-path solutions and if
alarm is enabled
40270
219
Path 2 fault disappears
Last fault
disappear
date
dd-mm-
yyyy,
hh:mm:ss
Read
Last time the fault disappeared.
Valid only for 2-path solutions and if
alarm is enabled
5
220
Dual slope alarm output
enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on
alarm output / call-up
40273
221
Dual slope fault hours
e.g. 0 h
0 to 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40274
222
Dual slope fault counter
e.g. 0
0 to 65535 Read
Total number of faults.
Valid only if alarm is enabled
40275
223
Dual slope fault appears Last fault
appear
date
dd-mm-
yyyy,
hh:mm:ss
Read
First time the fault appeared.
Valid only if alarm is enabled
40278
224
Dual slope fault disap‐
pears
Last fault
disappear
date
dd-mm-
yyyy,
hh:mm:ss
Read
Last time the fault disappeared.
Valid only if alarm is enabled
6
225
Parameter checksum
alarm output enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on
alarm output / call-up
40281
226
Parameter checksum
fault hours
e.g. 0 h
0 ... 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40282
227
Parameter checksum
fault counter
e.g. 0
0 ... 65535 Read
Total number of faults. Valid only if alarm
is enabled
40283
228
Parameter checksum
fault appears
Last fault
appear
date
dd-mm-
yyyy,
hh:mm:ss
Read
First time the fault appeared.
Valid only if alarm is enabled
40286
229
Parameter checksum
fault disappears
Last fault
disappear
date
dd-mm-
yyyy,
hh:mm:ss
Read
Last time the fault disappeared.
Valid only if alarm is enabled
7
230
Low power alarm output
enable
Yes
No, Yes
R/W
Select "Yes" to enable active alarm on
alarm output / call-up
40289
231
Low power fault hours
e.g. 0 h
0 to 65535 Read
Total hours fault active.
Valid only if alarm is enabled
40290
232
Low power fault counter
e.g. 0
0 to 65535 Read
Total number of faults.
Valid only if alarm is enabled
40291
233
Low power fault appears Last fault
appear
date
dd-mm-
yyyy,
hh:mm:ss
Read
First time the fault appeared.
Valid only if alarm is enabled
Parameter lists
A.3 Diagnostic
FUS380/FUE380
90