Modbus RTU RS-485
17
SFIDK.PS.023.D2.02
Modbus
Modbus
No. of
Data
Read/
register
address
bytes
type
Description
write
4:02101
2100
2
Byte
Totalizer 1 selection:
1 = Massflow
2 = Fraction A
R/W
3 = Fraction B
4 = Volumeflow
4:02102
2101
2
Byte
Totalizer 1 direction:
0 = Reverse
1 = Forward
R/W
2 = Net
4:02103
2102
2
Byte
Totalizer 2 selection:
1 = Massflow
2 = Fraction A
R/W
3 = Fraction B
4 = Volumeflow
4:02104
2103
2
Byte
Totalizer 2 direction:
0 = Reverse
1 = Forward
R/W
2 = Net
MASS 6000
6.
Modbus holding registers
Modbus
Modbus
No. of
Data
Read/
register
address
bytes
type
Description
write
4:02102
2101
2
Byte
Totalizer 1 direction:
0 = Reverse
1 = Forward
R/W
2 = Net
4:02104
2103
2
Byte
Totalizer 2 direction:
0 = Reverse
1 = Forward
R/W
2 = Net
MAG 6000
Modbus
Modbus
No. of
Data
Read/
register
address
bytes
type
Description
write
4:02201
2200
2
Byte
Current Output 1 selection:
0 = Off
R/W
1 = On
4:02202
2201
2
Byte
Current Output 1 direction:
1 = Unidirectional
R/W
2 = Bidirectional
4:02203
2202
2
Byte
Current Output 1 range:
0 = 0 - 20 mA
1 = 4 - 20 mA
R/W
2 = 4 - 20 mA + Alarm
4:02204
2203
4
Float
Current Output 1 time constants
R/W
4:02206
2205
2
Byte
Digital Output 1 function:
0 = Off
1 = Pulse
2 = Frequency
R/W
3 = Error Level
4 = Error Number
5 = Direction/Limit
6 = Batch
4:02208
2207
2
Byte
Pulse 1 direction:
1 = Unidirectional
R/W
2 = Bidirectional
MAG 6000
6.5 Totalizer
6.6 Output