Replacing boards / replacing components
6 - 17
Siemens AG
Register 5
RXR2-130.061.02
Page 17 of 18
SIREMOBIL Compact
Medical Engineering
Rev. 03
10.03
CS PS 24
System Manual
Multispot 2000
6
Before installing the new Multispot 2000, set the correct refresh rate.
Refer to Table 1.
•
Disconnect the SIREMOBIL Compact line voltage plug from the receptacle.
•
Replace the Multispot 2000.
•
The screws securing the Multispot 2000 are accessible on the underside of the camera.
•
Perform a camera function test after the replacement.
•
Generate and program the exposure parameters for the Multispot 2000 according to the
IQ Quick test (refer to the log book). Record the new exposure parameters in the IQ
Quick test instructions.
Area dose product measurement device
6
•
Remove the SIREPHOS cover (see Chapter 1, "Covers").
•
Replace the electronics board and the dose measurement chamber.
•
Perform the calibration of the area/dose product measurement device. Refer to Chapter
5 "Calibrating the Area Dose Product Measurement".
•
Test the accuracy of the area dose product measurement device. Refer to Chapter 5
"Testing the accuracy of the area dose product measurement device.
•
Re-attach the SIREPHOS cover.
•
Seal the SIREPHOS cover using sealing compound (see Chapter 1, "Covers").
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
50Hz refresh rate
60Hz refresh rate
Jumper
Position
Jumper
Position
X4
b
X4
a
X5
open
X5
open
X6
b
X6
a
X7
b
X7
a
X8
b
X8
a
X9
b
X9
a
X10
b
X10
a
X11
b
X11
a
Tab. 1
WARNING