SIREMOBIL Compact
Register 5
RXR2-130.061.02
Page 22 of 22
Siemens AG
System Manual
Rev. 03
10.03
CS PS 24
Medical Engineering
5 - 22
Adjustments / Programs
Memoskop, configuration for the number of existing monitors
5
When you replace the Memoskop, it will be configured for 2 monitors. This configuration
displays USER SETUP and TECHNICAL SETUP on the second monitor. Therefore,
USER SETUP and TECHNICAL SETUP are not visible for SIREMOBIL Compact systems
with only one monitor. To reconfigure the system to 1 monitor, proceed as follows:
•
Switch the SIREMOBIL Compact on.
•
On the Memoskop, reconnect the video cable from OUT 1 to OUT 2.
•
Press CTRL + T at the same time on the Memoskop keyboard;
TECHNICAL SETUP is called up and is now visible on the monitor.
•
Call up the menu "Number of monitors" by pressing the "3" key and subsequently the
"RETURN" key on the Memoskop keyboard.
•
With the -> and <- cursor keys, set the number to 1 monitor and press the "RETURN"
key.
The text display disappears on the monitor.
•
Press the "RETURN" key once more.
•
Reconnect the monitor to the OUT 1 video output.
The text display is now visible on the monitor.
•
Exit TECHNICAL SETUP by pressing the "HOME" key on the Memoskop keyboard.