SIREMOBIL Compact
Register 5
RXR2-130.061.02
Page 20 of 22
Siemens AG
System Manual
Rev. 03
10.03
CS PS 24
Medical Engineering
5 - 20
Adjustments / Programs
9.
Move text positions
Allows you to configure the position of superimposed text.
10. Date / Time format
European: TT MM JJ / HH MM
US:
MM DD YY / HH MM
11. System configuration
Allows you to read out the current Memoskop configura-
tion.
12. Keyboard
Allows you to program the corresponding keyboard layout
Universal / Roman / English.
13. Language
Allows you to configure the language.
14. Graphics level
Level (brightness) of superimposed text.
(Default: 128)
15. Display of real time clock Yes: Real time clock (date and time) are displayed
depending on the User setup programming for
Memoskop. The real time clock is active.
NO: The real time clock is switched off. The User setup
programming will not affect it.
16. Display of rotated images Yes: When pressing the image rotation key, the current
image will be computerized and displayed.
NO: The rotated image will not be computerized when
pressing the image rotation key. For SIREMOBIL Compact
always set to YES.
17. Monitor zoom
Yes: When pressing the key for Memoskop - Zoom, the
image will be displayed enlarged.
NO: The Memoskop - Zoom key has no function.
For SIREMOBIL Compact set on NO.
18. Monitor split
Yes: Horizontal or vertical image split.
No: No horizontal or vertical image split possible.
For SIREMOBIL Compact with 1 monitor:
program Yes.
For SIREMOBIL Compact with 2 monitors:
program No.
19. Monitor text
Yes: Superimposes text.
No: No text superimposition.
For SIREMOBIL Compact always set on Yes.
20. Maximum image numbers The number of stored images can be programmed.
Program the value to 700 for Memoskop C.
21. Product name
Allows you to change the Memoskop name, displayed
during system booting. Do not change this for SIREMOBIL
Compact !
22. Diagnostics
Allows you to invoke test routines for repair by the manu-
facturer of the Memoskop.
Function:
Comments: