Replacing boards / replacing components
6 - 15
Siemens AG
Register 5
RXR2-130.061.02
Page 15 of 18
SIREMOBIL Compact
Medical Engineering
Rev. 03
10.03
CS PS 24
System Manual
•
Perform the IQ quick test.
Memoskop C
6
•
Replace the Memoskop C.
The attachment screws are accessible on the underside of the storage device.
•
Set the Memoskop C to the correct video frequency and configure it for the number of
existing monitors.
Refer to Chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors".
•
Configure the following parameters in Technical Setup for Memoskop:
E Monitor split
Configure horizontally. Configure vertically only upon cus-
tomer request.
F Time
Configure Yes.
G Date / Time setting
Set the current date and time.
H Hospital name
Program the hospital name.
I Storing the dose
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise, configure No.
K Dose documentation
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Read and record the parameters for Memoskop C if you can
select User setup. After replacing the Memoskop, configure the
Memoskop C-E according to the parameters noted.
3 Number of monitors
Configure 1 for SIREMOBIL Compact with 1 monitor.
Configure 2 for SIREMOBIL Compact with 2 monitors.
4 Hardcopy setup
Check whether 5500 is programmed in "4 Max Hardcopy
time". If not, program this value.
5 Edge 1 coef set
Check whether 3 is programmed, If not, program this value.
6 Edge 2 coef set
Check whether 8 is programmed, If not, program this value.
7 Motion detection setup
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
10 Date / Time format
Set the country-specific format.
12 Keyboard
Set to the keyboard layout being used.
13 Language
Set for language of the country.
15 Display of realtime clock Configure Yes. Select No only upon customer request.
16 Display of rotated images Configure Yes.
17 Monitor zoom
Configure No.
18 Monitor split
Configure Yes for SIREMOBIL Compact with 1 monitor.
Configure No for SIREMOBIL Compact with 2 monitors.
19 Monitor text
Configure Yes.
20 Maximum image number Configure 100.
NOTE