Adjustments / Programs
5 - 19
Siemens AG
Register 5
RXR2-130.061.02
Page 19 of 22
SIREMOBIL Compact
Medical Engineering
Rev. 03
10.03
CS PS 24
System Manual
Overview of SIREMOBIL Compact parameters
5
Invoke the Moreinfo.hlp help file on the service PC for a short description of the parame-
ters available on the SIREMOBIL Compact.
Overview of Memoskop C / C-E parameters
5
User Setup
•
The User Setup is invoked with the CTRL + U key combination.
•
The available functions are described in the SIREMOBIL Compact Operating
Instructions.
Technical Setup
•
The technical setup is invoked with the CTRL + T key combination.
The following functions can be invoked:
If parameters have been changed and saved in the "Adjustments,
Parameter" menu with Save to file, you must exit the service pro-
gram with "Logoff" and "Quit" to save the modified parameters to
the disk.
Function:
Comments:
1.
Test pattern generation
Not for service, invoke the necessary test images on the
service PC.
2.
Video frequency
The currently programmed frame rate is shown in the
parentheses to the right. After downloading the
Memoskop or after replacement, the Memoskop will start
up at the default video frequency (50 Hz).
If the monitors are configured for a different video fre-
quency, you must perform "blind" configuration for the cor-
rect video frequency.
Refer to Configuring video frequency.
3.
Number of monitors
Allows you to program the number of monitors used (1 or
2) .
4.
Hardcopy setup
You can configure the connection for hardcopy cameras.
Check whether 5500 is programmed for "4 Max Hardcopy
time" and if not, program this value.
5.
Edge 1 coef set
Allows configuration of edge enhancement. Check
whether 3 is programmed and if not, program this value.
6.
Edge 2 coef set
Allows Configuration of edge enhancement. Check
whether 8 is programmed and if not, program this value.
7.
Motion detection setup
Configures the motion detector.
Check if at MD 1: Threshold 100 and at MD 2: Threshold
120 is programmed, if not, program these values.
8.
Circle mask setup
Allows you to configure the diameter of the circle mask.
NOTE