Replacing boards / replacing components
6 - 3
Siemens AG
Register 5
RXR2-130.061.02
Page 3 of 18
SIREMOBIL Compact
Medical Engineering
Rev. 03
10.03
CS PS 24
System Manual
•
Check the dose rate and adjust it if necessary.
•
Check the image sharpness with the IQ Quick test and, if necessary, adjust it at the I.I.
optics.
•
Perform the IQ Quick test.
Power supply M14, +5V / +15V / -15V
6
•
Replace the power supply.
•
Check the power supply voltage and adjust it if necessary. Refer to Chapter 3 of these
instructions and check the operating voltages.
Power supply M13, +12.3V Videomed DC
6
•
Replace the power supply.
•
Check the power supply voltage and adjust it if necessary. Refer to Chapter 3 of these
instructions and check the operating voltages.
I.I. mini voltage supply
6
•
Replace the control panel of the I.I. mini voltage supply (installed above Videomed DC.).
•
Replace the high voltage panel of the I.I. mini voltage supply (Mounted on the C-arm,
accessible after removing the cover near the I.I., on the outer edge of the C-arm. Refer to
Chapter 1, "Information regarding this document").
•
Refer to the I.I. test protocol for the voltages E1 / E2 / E3 and A. Check and adjust them, if
necessary, according to Chapter 3, "Checking the operating voltages".
•
Check the collimation for survey and zoom formats and adjust it if necessary.
•
Check the overall resolution according to the IQ Quick test.
•
Perform the IQ Quick test.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
WARNING
WARNING
WARNING